Correct treatment of sudden situation in the use of paper rupture tester

The paper rupture testing machine is divided into manual and fully automatic. No matter which one, it must be trained before the operation, or the operating instructions of the testing machine. If the operator does not go through the training, the operation is Certainly difficult, especially if you don't pay attention to some small details, there will be similar emergencies. For example, after breaking the rubber mold and the rubber mold is broken, the operator will mistakenly think that the equipment is faulty, of course not In this way, this is only caused by mistakes in the operation process.

Let us analyze the operation process of the rupture tester and the emergency treatment method:

Take the manual rupture machine; in the case of ensuring that there is no problem with glycerin and the whole equipment before the test, the first sample is placed in the pressure plate, and the upper pressure plate is manually pressed to press the upper pressure plate. The pointer on the pressure gauge is between seven and ten. After the sample is pressed:

After pressing the clear key, press the peak key, then grab the pressure rod with your right hand and put the left hand into the pressure plate sample to sense the crack of the sample. After both hands are placed in the specified position, the right hand will force the pressure rod toward Pull the left side, and then observe the change in the value on the display. At the same time, pay attention to whether the left hand touches the sample on the sample. If the sample is broken, the right hand should release the pressure rod, that is, push the pressure rod to the right. It takes two to three seconds to loosen the pressure rod at the moment of breaking the sample. If the pressure is not released, it is easy to break the rubber mold, so this part must pay attention.

1. What if the rubber mold is broken during the operation?

If there is a situation in which the rubber mold is broken, it is not as good as we think of replacing the rubber mold with new ones. This is wrong. Because when the rubber mold is broken, the glycerin system may enter the air, so the value of the test is not accurate, and the rubber mold is always concave. Normally, the rubber mold must be bulged, and the hand is pressed. It is recessed and returns to its original position when the hand is released.

2. How to deal with the air entering the glycerin? How can I discharge the air inside the glycerin system?

First start the test machine, grab the presser bar with your right hand, pressurize the test machine, and observe the degree of the rubber mold bulging. When the rubber mold is bulged to a certain height, that is, the rubber mold is bulged higher than the lower pressure plate. When the centimeters is high, the left hand presses the stop button, the right hand cannot release the pressure lever. After pressing the stop button, the left hand releases the screw on the oil cup, and the screw is taken out, then the start button is pressed and the right hand is released. When the equipment is venting air, it can also be seen that the glycerin in the oil cup is also falling.

3. If the rubber mold is not bulged after the above operation, what should we do?

If the rubber mold still fails to bulge after the operation is completed, repeat the operation as many times as above. However, please remember to turn the screw back when venting the air inside the glycerin, otherwise the glycerin in the oil cup will be discharged together.